What are you looking for?

Find tests, products, news and congresses across our website.

Your selection

Insert request

Complete request

Product information

Tecnoclot® PT Owren manual

Transferred from the original PDF and compared with it. The original document remains authoritative.

Coagulation assay for the quantitative determination of Prothrombin Time (PT) in human citrated plasma. The lyophilised reagent is sensitive to abnormal levels of coagulation factors II, VII and X and is used for the monitoring of oral anticoagulant therapy (i.e. warfarin). The reagent contains rabbit brain thromboplastin and adsorbed bovine plasma as a source for the coagulation factors V and fibrinogen and CaCl₂.

Technoclot® PT Owren replaces Thrombotest und Normotest.

Test principle

Prewarm the reagent at 37°C and add the sample. Start timing of the coagulation.

Reagents and stability (after reconstitution)

ReagentStability / note
Technoclot PT Owren manual1 hour at 37°C 6 hours at RT 24 hours at 2-8°C 1 month at -20°C
Pack sizeArticle number
10 x 4 ml5005032
10 x 10 ml5005037

Date: 29.02.2016