Fully automated coagulation analysers
The Ceveron® 100-Series comprises coagulation analysers that can detect the parameters of classical coagulation diagnostics as well as thrombin generation.
By using a quenching module, factor XIII activity and the ADAMTS-13 activity can be determined fully automatically with the Ceveron® s100.
The Ceveron® 100-Series includes
Ceveron® c100: classic haemostasis assays, chromogenic and turbidimetric assays |
Ceveron® t100: classic haemostasis assays, chromogenic and turbidimetric assays and additional thrombin generation (TGA) |
Ceveron® s100: classic haemostasis assays, chromogenic and turbidimetric assays and additional TGA, Factor XIII activity, and ADAMTS-13 activity. |
Specifications of the fully automated Ceveron®-Series
Methods | Clotting, chromogenic, turbidimetric assays, TGA module* and quenching module* |
Optic | 4 channels with 4 wave lengths each: 405, 540, 630, 740 nm 4 channels TGA module* 4 channels quenching module* |
Samples on board | 36 (24 Primary tubes +12 sample cups) |
Reagents on board | 32, allocated on 14 (12-18 °C), 10 (RT) 8 buffer and dilutions |
Cuvettes on board | 84 |
Tests per sample | 36 |
Sample volume | 5-400 µL |
Reagent volume | 5-400 µL |
Throughput | 75 PT/h |
Emergency samples | Yes, up to 8 samples |
Quality control | Integrated, user defined QC limits, Westguard rules |
Patient memory | unlimited |
Operating system | Windows® 10 |
Seize (device) | 750 mm x 675 mm x 420 mm (D x W x H) |
Weight | 40 kg |
High precision, flexibility and throughput
Rerun and reflex testing in combination with very low cv in routine and speciality testing, guarantee secure results for perfect diagnostics.
Having a minimal patient and reagent dead volume, adult´s and children´s samples can be run without any differentiation.
Due to the workflow of Ceveron® 100 series an exact incubation time on each measurement is guaranteed.
Easy access to all raw data, help comparing patient results, QC results and calibration curves.
- Westguard Rules
- Laboratory specific confidence range
- 3s range
- Lifetime of validation adjustable
With a throughput of 75 PT/hour, a TGA result below 20 minutes and an ADAMTS13 activity result within 25 minutes, Ceveron® 100 series instruments are the perfect solution for fast and precise diagnostic of patients.
Ceveron® 100 series is ready to use within 5 minutes after switching on the instrument.
No daily maintenance needed. Just 15 minutes once per week.
Handling of samples and reagents is completely traceable. This is indepedent from entering the samples via barcode or manually.
The reagent lots can be send to the LIS.
All temperatures during pipetting, incubation and measurement are displayed and stored internally.
Semi-automated coagulation analysers
The Ceveron® one, four and ten instruments comprises semi-automated coagulation analysers that use a rotating ball to detect the end point of coagulation. Ceveron® four and ten are available with an additional optical unit.
The Ceveron® m instruments with a unique detection principle and in combination with high-level analytical algorithm, measures exact, precise and reproducible results. Operation via intuitive, colored touchscreen, as well as patient result management are perfectly optimized.
The Ceveron® m1 has one channel, the Ceveron® m2 has 2 channels and the Ceveron® m1 has 4 channels
Ceveron® one, four and ten
Specifications of the Ceveron® one, four, ten series
Ceveron® one, four and ten are semi-automated coagulation analysers to detect coagulation and fibrinolysis.
Mechanical clot detection: a magnetic sensor monitors the rotation of a steel ball in the cuvette. Clot formation stops the rotation of the ball and this is recorded by the sensor as the test end point.
Specifications | Ceveron® one | Ceveron® four | Ceveron® ten |
Mechanical channels | 1 | 4 | 10 |
Optical channel (405/650 nm) | -- | 1 (optional) | 1 (optional) |
Incubation positions (37°C) | 2 | 4 | 10 |
Sample preparation positions (RT) | -- | 8 | 20 |
Reagent positions (37° C) | 2 | 5 | 10 |
Cuvette volume | 250 – 600 µL | 250 – 600 µL | 250 – 600 µL |
Temperature controlled (37° C) | -- | 2 | 2 |
Pre-defined tests | PT, INR | PT, aPTT, TZ, FIB, Factors | PT, aPTT, TZ, FIB, Factors |
Free definable tests | -- | AT, anti-Xa, D-Dimer, chromogenic assays… | AT, anti-Xa, D-Dimer, chromogenic assays… |
Result output | %, INR, ratio, Sec | %, INR, ratio, Sec., min, h, g/L, mOD, free definable | %, INR, ratio, Sec., min, h, g/L, mOD, free definable |
Calibration curve | up to 4 points | up to 5 points | up to 5 points |
Clotting tests - Capillary Blood/Venous Blood | yes | yes | yes |
Clotting tests – Plasma | yes | yes | yes |
Chromogenic tests | -- | yes (optional) | yes (optional) |
Barcode reader interface | -- | yes | yes |
Patient results memory | 100 | 1000 | 1000 |
External printer | yes | yes | yes |
Dimensions (D x W x H) | 218 x 125 x 93 mm | 430 x 420 x 17 mm | 430 x 590 x 170 mm |
Weight | 1,2 kg | 11 kg | 15 kg |
LIS-interface | -- | yes | yes |
PC-interface | -- | yes | yes |
The Ceveron® four and ten are available with an optional optical channel for chromogenic assays
Suitable for whole blood and plasma samples
Ceveron® m1, m2 and m4
Specifications of the semi-automated Ceveron® m1, m2 and m4
Ceveron® m1, m2 and m4 are semi-automated coagulation analysers to detect coagulation (PT Owren, PT, aPTT, FIB, TT). The measuring principle is based on the change of the cloudiness during the coagulation reaction.
Specifications | Ceveron® m1 | Ceveron® m2 | Ceveron® m4 |
Channels | 1 | 2 | 4 |
Wave length | 620 (red) | 405 (UV) | 405 (UV) |
Pre calibrated assays | yes | yes | yes |
Pre warm cuvettes | 10 | 20 | 20 |
Pre warm reagent positions 24 mm (piece) 22 mm (piece) 11 mm (piece) |
1 2 2 |
1 2 2 |
1 2 2 |
Reagent mixing position | - | 1 | 1 |
Sample material | Whole blood Plasma | Plasma | Plasma |
Parameters | PT Owren PT, aPTT, FIB, TT | PT Owren PT, aPTT, FIB, TT | PT Owren PT, aPTT, FIB, TT |
RS232 | 2 | 2 | 2 |
USB | 2 | 2 | 2 |
Temperature (Pre incubation positions) | 37°C | 37°C | 37°C |
Display | Touchscreen (4.3“/109 mm) | Touchscreen (4.3“/109 mm) | Touchscreen (4.3“/109 mm) |
Dimensions (device) | 230 x 148 x 94 mm (L x B x H) | 230 x 148 x 94 mm (L x B x H) | 230 x 148 x 94 mm (L x B x H) |
Power supply | 110 – 240 V (ac)/ 5 V (dc), 3,3 A | 110 – 240 V (ac)/ 5 V (dc), 3,3 A | 110 – 240 V (ac)/ 5 V (dc), 3,3 A |