Intended use
To control the photometer function of the 4 measuring channels and the 4 mixer motors of APACT 4004.
Contents
3 x 2 ml vials with different solutions, ready to use. Photometer-Control 1 = PPP Photometer-Control 2 = PRP Photometer-Control 3 = 50 % Solution
Procedure
APACT 4004
Select assay QC-Test on the APACT and LPC-Software
Cuvettes
Shake the Photometer-Controls very well before use. Pipette 200 µl of each Photometer –Control into a cuvette. Use the cuvette of Photometer-Control 1 for all 4 channels to store PPP-value. Use the cuvette of Photometer-Control 2 for all 4 channels to store PRP-value.
Use the cuvette of Photometer-Control 3 for all 4 channels to measure the 50 % value.
Expected values
40 – 60 % max. Aggregation The differences from the mean value of channel 1 -4 should not be higher than +/- 4 % max. Aggregation.
Example:
| Channel 1 | Channel 2 | Channel 3 | Channel 4 |
|---|---|---|---|
| 47,0 | 49,0 | 47,0 | 49,0 |
Mean % Aggreg. = 48,0 (= range 44,0 – 52,0 )
QC = okay, all values are in the range